Sign In to gain access to subscriptions and/or personal tools.
|
A new oxygenator change-out system and procedure
U Da Broi
Azienda Ospedaliera Santa Maria Della Misericordia Anaesthesiology Department, Udine, Italy
V Adami
E Falasca
W Malangone
S Crini
SB Technology, Udine, Italy
A Degrassi
DPMSC and Consorzio Fenice, University of Udine, Udine, Italy
Official reports relate that, in the US, one patient/month dies as a result of the emergency oxygenator change-out procedure, and the permanent injury of some patients is the result of current oxygenator change-out procedures or oxygenator failures, both in extracorporeal circulation (ECC) and extracorporeal membrane oxygenation (ECMO).
The aim of this article is to evaluate a new system and procedure, dedicated to oxygenator change-out, represented by two three-way stopcocks inserted in the ECC line in use. A dedicated back-up oxygenator and circuit can be easily primed and connected to the dedicated connector on the stopcocks, then blood flow is diverted to the new oxygenator without interruption of the ECC.
Tests performed showed that oxygenator change-out can be completed by perfusionists in 62.139+11.12 sec. Results obtained show that the new system and procedure allows fast, safe and reproducible oxygenator change-out without interruption of the ECC.
References
- Wahba A, Philip A, Behr R, Birnbaum DE. Heparincoated equipment reduces the risk of oxygenator failure . Ann Thorac Surg 1998; 65: 1310-1312 .[Abstract/Free Full Text]
- Niimi Y, Ichinose F, Ishiguro Y et al. The effects of heparin coating of oxygenator fibers on platelet adhesion and protein adsorption . Anesth Analg 1999; 89(3): 573-579 .[Abstract/Free Full Text]
- Usui A, Hiruora M, Kawamura M. Heparin coating extends the durability of oxygenators used for cardiopulmonary support . Artif Organs 1999; 23(9): 840-844 .[CrossRef][Web of Science][Medline]
[Order article via Infotrieve]
- Wendel HP, Philipp A, Weber N, Birnbaum DE, Ziemer G. Oxygenator thrombosis: worst case after development of an abnormal pressure gradient-incidence and pathway . Perfusion 2001; 16(4): 271-278 .[Abstract/Free Full Text]
- Schaadt J. Oxygenator thrombosis . Perfusion 1999; 14(6): 425-435 .[Abstract/Free Full Text]
- Ginther R Jr, Fillingham R, Searles B, Darling E. Departmental use of perfusion crisis management drills: 2002 survey results . Perfusion 2003; 18(5): 299-302 .[Abstract/Free Full Text]
- Fisher AR, Baker M, Buffin M et al. Normal and abnormal trans-oxygenator pressure gradients during cardiopulmonary bypass . Perfusion 2003; 18(1): 25-30 .[Abstract/Free Full Text]
- Schabel RK. Ten common perfusion problems: prevention and treatment protocols . J Extra-Corpor Technol 1987; 19: 392-398 .
- Ginther R Jr., Fillingham R, Searles B, Darling E. Departmental use of perfusion crisis management drills: 2002 survey results . Perfusion 2003; 18(5): 299-302 .[Abstract/Free Full Text]
- Groom RC, Forest RJ, Cormack JE, Niimi KS, Morton J. Parallel replacement of the oxygenator that is not transferring oxygen: the PRONTO procedure . Perfusion 2002: 17(6): 447-450 .[Abstract/Free Full Text]
Perfusion, Vol. 21, No. 5,
297-303 (2006)
DOI: 10.1177/0267659106074771

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
J. B. Riley, P. D. Scott, and G. J. Schears
Update on Safety Equipment for Extracorporeal Life Support (ECLS) Circuits
Seminars in Cardiothoracic and Vascular Anesthesia,
September 1, 2009;
13(3):
138 - 145.
[Abstract]
[PDF]
|
 |
|
|
|