Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Sign In to gain access to subscriptions and/or personal tools.
Perfusion
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Fried, D. W.
Right arrow Articles by Zombolas, T. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Fried, D. W.
Right arrow Articles by Zombolas, T. L.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati   Add to Twitter  
What's this?

Clinical oxygen transfer performance of the Sorin Monolyth membrane oxygenator

David W. Fried

Albert Einstein Medical Center, Philadelphia, Pennsylvania

Benjamin N. DeBenedetto

Albert Einstein Medical Center, Philadelphia, Pennsylvania

Joseph J. Leo

Albert Einstein Medical Center, Philadelphia, Pennsylvania

Gabriel J. Mattioni

Albert Einstein Medical Center, Philadelphia, Pennsylvania

Hasratt Mohamed

Albert Einstein Medical Center, Philadelphia, Pennsylvania

Theodore L. Zombolas

Albert Einstein Medical Center, Philadelphia, Pennsylvania

Oxygen transfer performance of the Sorin Monolyth membrane oxygenator was evaluated. Similar to previous oxygen transfer performance studies conducted at this institution, our purpose was sixfold: (1) to construct an oxygen transfer slope (OTS); (2) to find the maximum extrapolated oxygen transfer; (3) to calculate the oxygenator performance index (OPI); (4) to generate a shunt fraction line; (5) to determine the percentage of predicted shunt (POPS); and (6) to compare the Monolyth's performance to several previously studied membrane oxygenators. From the OTS, the maximum extrapolated oxygen transfer was 346.4 ml O2/min. This absolute value was the lowest of the four oxygenators compared. When maximum oxygen transfer was compared relative to membrane surface area, the Monolyth ranked third (157.5 ml O2/Min). The Monolyth produced a relatively narrow range for the OPI (81.64-130.47%) and had the lowest standard deviation (SD) in this group. The Monolyth exhibited higher shunt fractions over the range of clinical blood flows when compared to our three previously studied oxygenators. The range of POPS values (71.65-128.77%) was relatively narrow and the SD was the lowest of the four. We concluded from our evaluation that the Monolyth had relatively low top end oxygen transfer capabilities, but provided very consistent and predictable oxygen transfer performance.

Perfusion, Vol. 9, No. 2, 119-126 (1994)
DOI: 10.1177/026765919400900206


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter    What's this?


This article has been cited by other articles:


Home page
PerfusionHome page
A. H Stammers, L. W Fristoe, A. Alonso, Z. Song, and T. Galbraith
Clinical evaluation of a new generation membrane oxygenator: a prospective randomized study
Perfusion, May 1, 1998; 13(3): 165 - 175.
[Abstract] [PDF]


Home page
PerfusionHome page
D. W Fried, T. L Zombolas, J. J Leo, H. Mohamed, and G. J Mattioni
Clinical oxygen transfer comparison of the Terumo Capiox SX18 and SX25 membrane oxygenators
Perfusion, March 1, 1998; 13(2): 119 - 127.
[Abstract] [PDF]


Home page
PerfusionHome page
D. W Fried, B. N DeBenedetto, and J. J Leo
Rethinking the AAMI/ISO 'International Standard' for oxygen transfer performance of artificial lungs
Perfusion, January 1, 1994; 9(5): 335 - 342.
[PDF]


Home page
PerfusionHome page
D. W Fried, B. N DeBenedetto, T. L Zombolas, and J. J Leo
Clinical evaluation of the Medtronic Maxima Plus membrane oxygenator
Perfusion, January 1, 1994; 9(5): 363 - 372.
[Abstract] [PDF]