Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

CiteULike is a free service for managing and discovering scholarly references - click here to get started.

Sign In to gain access to subscriptions and/or personal tools.
Perfusion
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Zacour, R. K
Right arrow Search for Related Content
PubMed
Right arrow Articles by Zacour, R. K
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati   Add to Twitter  
What's this?

Clinical evaluation of the Sarns membrane oxygenator

Richard K Zacour

Evanston Hospital, Illinois

A new membrane oxygenator, developed by Travenol Laboratories, has been introduced by Sarns/3M. The oxygenator is constructed of microporous polypropylene hollow fibres wound into a gas transfer module with an effective surface area of 1.8m2 and a total priming volume of 320ml. This was made possible by using computer-aided design techniques. Blood flow is perpendicular over the fibre in the module while the gas flow is directed within the fibres. The unit contains an integral stainless steel heat exchanger mounted beneath the fibre module. A selection of collapsible venous reservoirs is available to fulfil various patient requirements. The oxygenator and venous reservoir bag are mounted on a wire frame holder that can be arranged in any direction on the heart lung machine. The circuit was primed with 1800ml of a balanced electrolyte solution, 200ml of 25% albumin and 200ml of 25% mannitol, to which 4000 units of beef lung heparin were added. The 1000ml reservoir was used in all cases.

Seventeen consenting adult patients undergoing elective cardiac surgery participated in the clinical evaluation. Haematology and organ function studies indicated no significant changes attributable to the oxygenator. The mean postbypass plasma-free haemoglobin elevation was 8.2mg/dl. The mean postbypass haematocrit and platelet count reduction were 21 % and 27% respectively, indicating that platelet count tended to follow haemodilution with minimal loss. There were no gas exchange nor acid-base problems encountered. The mean PaO2 and PaCO 2 were 267mmHg and 42mmHg respectively with a maximum O2 transfer of 241 ml/min during the study.

Perfusion, Vol. 3, No. 3, 219-224 (1988)
DOI: 10.1177/026765918800300308


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter    What's this?


This article has been cited by other articles:


Home page
PerfusionHome page
D. W Fried and J. Bell-Thomson
Oxygen transfer performance of a new Sarns membrane oxygenator
Perfusion, October 1, 1992; 7(4): 263 - 271.
[Abstract] [PDF]


Home page
PerfusionHome page
D. W Fried and J. Bell-Thomson
Oxygen transfer efficiency of three microporous polypropylene membrane oxygenators
Perfusion, April 1, 1991; 6(2): 105 - 114.
[Abstract] [PDF]


Home page
PerfusionHome page
T. Gourlay, J. Fleming, K. Taylor, and M. Aslam
Evaluation of a range of extracorporeal membrane oxygenators
Perfusion, April 1, 1990; 5(2): 117 - 133.
[PDF]