Perfusion

 

Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Sign In to gain access to subscriptions and/or personal tools.
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Giavarina, D.
Right arrow Articles by Soffiati, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Giavarina, D.
Right arrow Articles by Soffiati, G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Perfusion, Vol. 17, No. 1, 23-26 (2002)
DOI: 10.1191/0267659102pf530oa

Monitoring high-dose heparin levels by ACT and HMT during extracorporeal circulation: diagnostic accuracy of three compact monitors

Davide Giavarina

Clinical Chemistry and Haematology Laboratory, Vicenza General Hospital, Vicenza, Italy

Mariarosa Carta

Clinical Chemistry and Haematology Laboratory, Vicenza General Hospital, Vicenza, Italy

Alessandro Fabbri

Department of Cardiovascular Surgery, Vicenza General Hospital, Vicenza, Italy

Jacopo Manfredi

Department of Cardiovascular Surgery, Vicenza General Hospital, Vicenza, Italy

Edoardo Gasparotto

Department of Cardiovascular Surgery, Vicenza General Hospital, Vicenza, Italy

Giuliano Soffiati

Clinical Chemistry and Haematology Laboratory, Vicenza General Hospital, Vicenza, Italy

The correct monitoring of heparin therapy and its reversal determines the successful conduct of cardiovascular surgery with extracorporeal circulation (ECC). The activated coagulation time (ACT) and the heparin management test (HMT) are the most frequently used tests in the operating room. Three compact monitors for ACT or HMT are here evaluated. Forty samples were obtained, at 10-min intervals, from eight patients during ECC. The ACT or HMT was immediately performed using: Hemochron JuniorTM ACT, CoaguCeckTM Pro (ACT) and Rapid Point Coag (HMT). Data were compared between them and with the heparin levels, measured as anti-Xa. The simple least squares linear regression among, respectively, Hemochron Junior ACT, CoaguCeck Pro ACT, Rapid Point Coag HMT and anti-Xa activity were i= 452.3, s= 15.2, Sy/x= 37.5, r= 0.18; i= 411.9, s= 22.1, Sy/x= 48.7, r= 0.21 and i= 479.4, s= 9.0, Sy/x= 9.3; r= 0.41. CoaguCeck Pro ACT results were above the upper detection limit (500 s) in 37 of 40 determinations. The comparison between ACT Hemocron and HMT Rapid Point Coag shows i= 35.7, s= 0.9, Sy/x= 35.4, r= 0.68, with a bias of 29.0 s (CI: 17.9-40.1), 95% of agreement between -41.5 s (CI: -60.7 to -22.3) and 99.5 s (CI: 80.4-118.7). Taking a concentration of 2.0 U/ml of heparin to discriminate between high- and low-risk conditions, receiver-operator characteristic (ROC) curve was used to rank the performance of the methods. Areas under the ROC curve± SE for Hemochron Junior ACT and Rapid Point Coag HMT were 0.629± 0.097 and 0.543± 0.096. The results obtained by HMT appear similar to those obtained by the ACT for monitoring high-dose heparin therapy in patients undergoing ECC. HMT appeared to perform better than ACT in measuring the heparin effect, while the ROC analysis gives a little more accuracy for ACT. Neither of the two methods is able to achieve enough evidence of diagnostic accuracy. Since these tests are widely used, and there are no laboratory alternatives, a real comparison with the outcome of the patients should be helpful for an evidence-based evaluation of these point-of-care tests.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?