Sign In to gain access to subscriptions and/or personal tools.
|
MAST system: a new condensed cardiopulmonary bypass circuit for adult cardiac surgery
K McCusker
Departments of Cardiothoracic Surgery, Portsmouth Regional Hospital, New Hampshire, kmccusker{at}heart-and-lung-center.com
V Vijay
Albert Einstein College of Medicine, Bronx
W DeBois
New York Presbyterian Hospital-Weill Cornell Center, New York
R Helm
Departments of Cardiothoracic Surgery, Portsmouth Regional Hospital, New Hampshire
D Sisto
Departments of Cardiothoracic Surgery, Portsmouth Regional Hospital, New Hampshire
There have been many refinements in cardiopulmonary bypass (CPB) techniques over the past few decades specific to design, materials and function. Despite these improvements, use of the standard length circuit tubing and pump oxygenator alter cellular, biochemical and rheological properties by inducing a systemic inflammatory response, persisting well into the early postoperative phase.
We have designed a new condensed CPB circuit, the MAST system, where the oxygenator and the pumps are brought closer to the operating table (within 30 inches) with the help of a series of telescopic swivel steel poles to which they are attached. The control console is retained at the usual remote location of 2 ft behind the MAST system. This configuration accomplishes a decrease in tubing length, priming volume and blood circulatory time within the extracorporeal circuit. Early experience of a hundred consecutive cases utilizing the MAST CPB system is presented along with a comparative analysis of prime volume, hemodilution and transfusion parameters of MAST system vs the low prime system, which is another newly developed CPB circuit utilizing a pediatric oxygenator to reduce prime volume and hemodilution.
Perfusion, Vol. 16, No. 6,
447-452 (2001)
DOI: 10.1177/026765910101600603

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
S. Gunaydin, T. Sari, K. McCusker, U. Schonrock, and Y. Zorlutuna
Clinical evaluation of minimized extracorporeal circulation in high-risk coronary revascularization: impact on air handling, inflammation, hemodilution and myocardial function
Perfusion,
May 1, 2009;
24(3):
153 - 162.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Lauten, K. Liebing, U. Franke, and T. Wahlers
The Jena universal perfusion system: a universal cardiopulmonary bypass circuit for cardiac surgery
Interactive CardioVascular and Thoracic Surgery,
February 1, 2007;
6(1):
1 - 4.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. G. Shann, D. S. Likosky, J. M. Murkin, R. A. Baker, Y. R. Baribeau, G. R. DeFoe, T. A. Dickinson, T. J. Gardner, H. P. Grocott, G. T. O'Connor, et al.
An evidence-based review of the practice of cardiopulmonary bypass in adults: A focus on neurologic injury, glycemic control, hemodilution, and the inflammatory response.
J. Thorac. Cardiovasc. Surg.,
August 1, 2006;
132(2):
283 - 290.e3.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. McCusker, A. Chalafant, G. de Foe, S. Gunaydin, and V. Vijay
Influence of hematocrit and pump prime on cerebral oxygen saturation in on-pump coronary revascularization
Perfusion,
May 1, 2006;
21(3):
149 - 155.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
E. J Fransen, Y. M Ganushchak, V. Vijay, D. S de Jong, W. A Buurman, and J. G Maessen
Evaluation of a new condensed extra-corporeal circuit for cardiac surgery: a prospective randomized clinical pilot study
Perfusion,
March 1, 2005;
20(2):
91 - 99.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Takai, K. Eishi, S. Yamachika, S. Hazama, T. Ariyoshi, and K. Nishi
Demonstration and Operative Influence of Low Prime Volume Closed Pump
Asian Cardiovasc Thorac Ann,
March 1, 2005;
13(1):
65 - 69.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Fiorucci, P. S. Gerometta, M. DeVecchi, C. Guzman, M. L. Costantino, and V. Arena
In vitro assessment of the vacuum-assisted venous drainage (VAVD) system: risks and benefits
Perfusion,
March 1, 2004;
19(2):
113 - 117.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
T Gourlay and P Connolly
Does cardiopulmonary bypass still represent a good investment? The biomaterials perspective
Perfusion,
July 1, 2003;
18(4):
225 - 231.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Gourlay, I. Samartzis, and K. M Taylor
The effect of haemodilution on blood/biomaterial contact-mediated CD11b expression on neutrophils: ex vivo studies
Perfusion,
March 1, 2003;
18(2):
87 - 93.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
V. Vijay and K. McCusker
Recent advances in biocompatible surface-modifying additives for cardiopulmonary bypass
Perfusion,
January 1, 2003;
18(1_suppl):
41 - 45.
[PDF]
|
 |
|
|
|