Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Click here to sign up for SAGE Journal Email Alerts today!

Sign In to gain access to subscriptions and/or personal tools.
Perfusion
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Segers, P A.
Right arrow Articles by Eilander, S
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Segers, P A.
Right arrow Articles by Eilander, S
Right arrowPubmed/NCBI databases
*Substance via MeSH
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati   Add to Twitter  
What's this?

Clinical evaluation of nine hollow-fibre membrane oxygenators

P AM Segers

Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle, peter.segers{at}isala.nl

J F Heida

Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle

I de Vries

Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle

C Maas

Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle

A J Boogaart

Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle

S Eilander

Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle

In a comparative study we investigated the performance characteristics of nine hollow-fibre oxygenators. In a clinical setting, 10 units of each type of oxygenator were tested for oxygen exchange, transoxygenator pressure drop, heat exchanger performance and blood trauma. The oxygenators included are Maxima PRF Plus, Affinity, Forté, Affinity NT, Quantum, Optima, Capiox 1.8, Hilite and Quadrox. Ninety patients scheduled for elective coronary bypass surgery were enrolled in the study. Cardiotomy suction blood was not allowed to mix with the circulating blood. Samples were taken preoperatively (t1), immediately after cooling and crossclamping of the aorta (t2), during rewarming (t3), and postoperatively (t4). The results showed that under clinical conditions all of the tested oxygenators are well capable of meeting the normal and above-normal oxygen needs. The Quantum and Affinity NT are the best performing oxygenators with respect to oxygen transfer. The Quantum oxygenator with a small membrane surface area of 1.9 m2 has also the highest oxygen transfer per square metre. A direct relationship between shunt fraction and blood flow could not be established in all oxygenators. The lowest shunt fraction at 4 litres blood flow was seen in the Maxima, Quantum and Affinity NT groups (15%). At near physiological PaO2-values, we noticed a negative influence on the shunt fraction. Large differences exist in pressure drop between the oxygenators under standardized conditions with circulating priming, as well as during bypass. The lowest pressure drop was noticed in the Quadrox and Affinity NT, the highest in the Quantum, Capiox and Optima. However, we did not measure any differences in pressure drop after cooling and during re-warming. Differences exist between the heat exchanger performances. The Quadrox oxygenator is the most efficient, while the Maxima has the smallest performance factor. It was not possible to calculate heat exchanger performance for the Hilite due to incorrect temperature measurement. The high pressure gradients in some of the devices did not result in higher haemolysis. Plasma free haemoglobin values were low in each oxygenator. There are no differences in platelet drop postoperatively. The influence on blood trauma of the higher pressure drop in some of the tested devices, in combination with the higher centrifugal pump revolutions needed to overcome this gradient, has to be studied with longer perfusion times.

Perfusion, Vol. 16, No. 2, 95-106 (2001)
DOI: 10.1177/026765910101600203


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter    What's this?


This article has been cited by other articles:


Home page
PerfusionHome page
A. Thiara, V Hoyland, H Norum, T. Aasmundstad, H. Karlsen, A. Fiane, and O Geiran
Extracorporeal membrane oxygenation support for 59 days without changing the ECMO circuit: a case of Legionella pneumonia
Perfusion, January 1, 2009; 24(1): 45 - 47.
[Abstract] [PDF]


Home page
PerfusionHome page
H. Dunningham, C. Borland, F. Bottrill, A. Vuylsteke, and D. Gordon
Modelling lung and tissue diffusion using a membrane oxygenator circuit
Perfusion, July 1, 2007; 22(4): 231 - 238.
[Abstract] [PDF]


Home page
PerfusionHome page
S Horton, C Thuys, M Bennett, S Augustin, M Rosenberg, and C Brizard
Experience with the Jostra Rotaflow and QuadroxD oxygenator for ECMO
Perfusion, January 1, 2004; 19(1): 17 - 23.
[Abstract] [PDF]