Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

CiteULike is a free service for managing and discovering scholarly references - click here to get started.

Sign In to gain access to subscriptions and/or personal tools.
Perfusion
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Chukwuemeka, A O
Right arrow Articles by Chambers, D J
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chukwuemeka, A O
Right arrow Articles by Chambers, D J
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati   Add to Twitter  
What's this?

A clinical evaluation of platelet function, haemolysis and oxygen transfer during cardiopulmonary bypass comparing the Quantum HF-6700 to the HF-5700 hollow fibre membrane oxygenator

A O Chukwuemeka

Department of Cardiothoracic Surgery, St Thomas’ Hospital, London, andrew.chukwuemeka{at}kcl.ac.uk

M RJ Turtle

Department of Cardiothoracic Surgery, St Thomas’ Hospital, London

U H Trivedi

Department of Cardiothoracic Surgery, St Thomas’ Hospital, London

G E Venn

Department of Cardiothoracic Surgery, St Thomas’ Hospital, London

D J Chambers

Department of Cardiac Surgical Research, St Thomas’ Hospital, London

The continued improvement of oxygenators is an important aspect of patient safety during cardiopulmonary bypass (CPB). The purpose of this study was to compare the Bard William Harvey HF-5700 oxygenator to the upgraded Bard Quantum HF-6700, which has recently been introduced into clinical practice. No clinical evaluation of this device has been published to date. The two oxygenators differ principally in that the Quantum has a smaller priming volume, achieved at the expense of a smaller membrane surface area which could result in sub-optimal gas exchange characteristics, increased haemolysis and increased platelet dysfunction during CPB.

Twenty adult patients undergoing elective, first time coronary artery bypass grafting (CABG) were randomly assigned either to the HF-5700 (n=10) or to the HF-6700 (n=10) group. One patient underwent mitral valve repair in addition to CABG and was excluded from further study. There were no statistically significant differences in either preoperative or operative parameters between the two groups. Samples were obtained at the start of CPB, at 30 min, 60 min, at the end of CPB and at 1 h following termination of CPB.

No significant differences between the two groups were found in oxygen transfer, haemolysis (plasma haptoglobin levels) or platelet function (a novel platelet activating factor (PAF)-induced platelet activation test) at any of the time points during CPB.

It was concluded that the Quantum HF-6700 matches the HF-5700 for the parameters studied, whilst having the advantage of requiring a smaller priming volume.

Perfusion, Vol. 15, No. 6, 479-484 (2000)
DOI: 10.1177/026765910001500602


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter    What's this?


This article has been cited by other articles:


Home page
Am. J. Respir. Crit. Care Med.Home page
A. Anthi, R. F. Machado, M. L. Jison, A. M. Taveira-DaSilva, L. J. Rubin, L. Hunter, C. J. Hunter, W. Coles, J. Nichols, N. A. Avila, et al.
Hemodynamic and Functional Assessment of Patients with Sickle Cell Disease and Pulmonary Hypertension
Am. J. Respir. Crit. Care Med., June 15, 2007; 175(12): 1272 - 1279.
[Abstract] [Full Text] [PDF]