Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Sign In to gain access to subscriptions and/or personal tools.
Perfusion
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Jaggy, C
Right arrow Articles by Turina, M
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jaggy, C
Right arrow Articles by Turina, M
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Affinity pump system: a new peristaltic blood pump for cardiopulmonary bypass

C Jaggy

Clinic for Cardiovascular Surgery, University Hospital, Zurich

M Lachat

Clinic for Cardiovascular Surgery, University Hospital, Zurich, lachat{at}chi.usz.ch

B Leskosek

Clinic for Cardiovascular Surgery, University Hospital, Zurich

G Zünd

Clinic for Cardiovascular Surgery, University Hospital, Zurich

M Turina

Clinic for Cardiovascular Surgery, University Hospital, Zurich

An in vitro study has been carried out to assess the pump performance of a new peristaltic, extracorporeal displacement pump (Affinity) for cardiopulmonary bypass. The pump system consists of a pump rotor (0-110 rpm), a pump chamber, a venous reservoir with a 5/8'' connecting tube and the Affinity console. The polyurethane chamber is connected to the venous reservoir by a 5/8'' tube and fills passively due to the hydrostatic pressure exhibited by the fluid height in the venous reservoir. The implementation of an occlusive segment in the pump chamber, which collapses in low filling states, should prevent significant negative pressures. An in vitro circuit was filled with bovine blood (37°C, hematocrit 35%) and the pump flow was measured by an ultrasonic transit time flow probe with respect to pre-load, diameter and length of attached tubing in the venous line, pump speed (rpm) and size of the connecting tube (3/8'' and 5/8'').

At 108 rpm and a preload equal to 10 mmHg, the flow was 8.6 ± 0.42 l/min for an afterload of 80 mmHg. The reduction of the inlet connector to 3/8'' diminished the pump flow significantly to 5.2 ± 0.31 l/min (p < 0.0001). The pump flow decreased linearly with respect to the length of the attached tube in the venous line and for a 2 m long 5/8'' silicon tube, the rpm-optimized flow was still 6.0 ± 0.28 l/min at a preload of 10 mmHg. In case of low filling state or too high rpm, the occlusive segment collapsed and no cavitation bubbles could be detected.

Our in vitro measurements yield a nomogram for rpm-optimized blood flow with respect to the pre-load in the venous reservoir. The delivered 5/8'' connecting tube facilitates optimum filling of the pump chamber for high blood flow, but limits the use of venous reservoirs to Affinity products. The pump yields a high blood flow even when long tubing in the venous line is used. This makes the pump a candidate for a ventricular assist device. In hypovolemia or high rpm, the occlusive segment collapses and no negative pressure is generated at the inflow site of the pump chamber.

Perfusion, Vol. 15, No. 1, 77-83 (2000)
DOI: 10.1177/026765910001500111