Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Sign In to gain access to subscriptions and/or personal tools.
Perfusion
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Taft, K. J
Right arrow Articles by Beck, D. J
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Taft, K. J
Right arrow Articles by Beck, D. J
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati   Add to Twitter  
What's this?

Cardioplegia flow dynamics in an in vitro model

Kimberly J Taft

Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha, Nebraska

Alfred H Stammers

Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha, Nebraska

Clinton C Jones

Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha, Nebraska

Melinda S Dickes

Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha, Nebraska

Michelle L Pierce

Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha, Nebraska

Daniel J Beck

Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha, Nebraska

The flow of fluids in extracorporeal circuits does not conform to conventional Poiseuille mechanics which confounds calculating cardioplegia (CP) flow distribution. The purpose of this study was to quantify CP flow dynamics in a model simulating coronary atherosclerosis across varying sized restrictions.

An in vitro preparation was designed to assess hydraulic fluid movement across paired restrictions of 51, 81 and 98% lumen reductions. Volume data were obtained at variable flow, temperature, viscosity and pressure conditions.

CP delivered through 14- and 18-gauge (GA) conduits at 8°C and 100 mmHg infusion pressure revealed that both four to one and crystalloid CP solutions had significantly less total percentage flow through the 14-GA conduit, p < 0.0001 and p < 0.001, respectively. Overall, 4:1 CP exhibited the most favorable fluid dynamics at 8°C in that it delivered the highest percentages of total CP flow through the smaller lumen conduit. At both 8°C and 37°C delivery, blood CP resulted in the least homogeneous fluid distribution at all delivery parameters. The results in relation to blood viscosity indicate that, although the 8°C blood CP had a significantly greater viscosity than 37°C blood CP, it did not produce an effect in fluid distribution.

These data show that increasing the cardioplegic solution hematocrit causes an inhomogeneous fluid distribution regardless of delivery temperature or infusion pressure.

Perfusion, Vol. 14, No. 5, 341-349 (1999)
DOI: 10.1177/026765919901400505