Sign In to gain access to subscriptions and/or personal tools.
|
The incidence and cause of emergency oxygenator changeovers
A R Fisher
Heart and Lung Department, Harefield Hospital, Harefield, Middlesex
Emergency oxygenator changeover presents as an important potential hazard during cardiopulmonary bypass. In order to assess the frequency and possible causes of this event a retrospective survey was conducted of cardiothoracic units in the UK and Ireland. The survey was sent to these units in 1993 to cover the period from 1990 to 1992 inclusive and again in 1997 to cover the period from 1994 to 1996 inclusive. The total number of procedures reported in 1993 was 68 937 (response of 63%), among which there were 17 emergency oxygenator changeovers representing an incidence of one for every 4055 cases. The use of aprotinin was associated with nine of these incidents. Two other incidents occurred due to leaking oxygenators. The total number of procedures reported in 1997 was 97 313 (response of 70%), among which there were 21 emergency oxygenator changeovers, representing an incidence of one for every 4634 cases. The use of aprotinin was only associated with two of these incidents while seven were due to the development of a high trans-oxygenator pressure gradient and six occurred due to leaking oxygenators.
Perfusion, Vol. 14, No. 3,
207-212 (1999)
DOI: 10.1177/026765919901400309

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
J. W. Hammon
Extracorporeal Circulation: Perfusion System
Card. Surg. Adult,
January 1, 2008;
3(2008):
350 - 370.
[Full Text]
|
 |
|

|
 |

|
 |
 
D. A Palanzo
Perfusion safety: defining the problem
Perfusion,
July 1, 2005;
20(4):
195 - 203.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
J Dubois, L Jamaer, U Mees, J-L Pauwels, F Briers, J Lehaen, and M Hendrikx
Ex vivo evaluation of a new neonatal/infant oxygenator: comparison of the Terumo CAPIOX(R) Baby RX with Dideco Lilliput 1 and Polystan Safe Micro in the piglet model
Perfusion,
September 1, 2004;
19(5):
315 - 321.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Ginther Jr, R. Fillingham, B. Searles, and E. Darling
Departmental use of perfusion crisis management drills: 2002 survey results
Perfusion,
September 1, 2003;
18(5):
299 - 302.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Noora, A. Lamy, K. M Smith, R. Kent, D. Batt, J. Fedoryshyn, and X. Wang
The effect of oxygenator membranes on blood: a comparison of two oxygenators in open-heart surgery
Perfusion,
September 1, 2003;
18(5):
313 - 320.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
A. R. Fisher, M. Baker, M. Buffin, P. Campbell, S. Hansbro, S. Kennington, A. Lilley, and M. Whitehorne
Normal and abnormal trans-oxygenator pressure gradients during cardiopulmonary bypass
Perfusion,
January 1, 2003;
18(1):
25 - 30.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
E. A. Hessel II and L. H. Edmunds Jr.
Extracorporeal Circulation: Perfusion Systems
Card. Surg. Adult,
January 1, 2003;
2(2003):
317 - 338.
[Full Text]
|
 |
|

|
 |

|
 |
 
R C Groom, R J Forest, J E Cormack, K S Niimi, and J Morton
Parallel replacement of the oxygenator that is not transferring oxygen: the PRONTO procedure
Perfusion,
December 1, 2002;
17(6):
447 - 450.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
A. H Stammers and B. L Mejak
An update on perfusion safety: does the type of perfusion practice affect the rate of incidents related to cardiopulmonary bypass?
Perfusion,
May 1, 2001;
16(3):
189 - 198.
[Abstract]
[PDF]
|
 |
|
|
|