Sign In to gain access to subscriptions and/or personal tools.
|
Clinical evaluation of a leucocyte-depleting blood cardioplegia filter (BC1B) for elective open-heart surgery
Ichiro Suzuki
Department of Clinical Engineering, Sendai Medical Centre, Sendai
Noboru Ogoshi
Department of Clinical Engineering, Sendai Medical Centre, Sendai
Miki Chiba
Department of Clinical Engineering, Sendai Medical Centre, Sendai
Tsunehiro Komatsu
Department of Cardiovascular Surgery, Sendai Medical Centre, Sendai
Yoshimasa Moizumi
Department of Cardiovascular Surgery, Sendai Medical Centre, Sendai
Activation of leucocytes during extracorporeal circulation has attracted attention in recent years as a cause of reperfusion injury in open-heart surgery patients. In the present study, 40 adult patients undergoing elective open-heart surgery were randomized into two groups: 20 using the Pall BC1B leucocyte-depleting filter for blood cardioplegia (group 1) and the other 20 without the filter (group 2). In order to determine if the filter was effective in protecting the myocardium, CPK-MB and troponin-T (TnT) were measured. In addition, efforts were also made to determine appropriate sites at which the BC1B blood cardioplegia filter should be positioned. There were no significant differences between the two groups in terms of conditions of perfusion. No adverse effects were seen in either group. The total leucocyte reduction rate through the filter was 98.1% with the passage of 2 litres of blood through the filter. A pressure drop of 4.4 ± 3.2 mmHg was observed through the filter during use. Statistically significant differences were noted between the two groups in CPK-MB (p = 0.031) and TnT (p = 0.004). Findings obtained in previous studies demonstrate that the various advantages of leucocyte reduction, shown in experimental studies, can be translated into clinical advantages. In conclusion, based on the results in this clinical study showing significant difference in CPK-MB and TnT which are known effective indicators for myocardial injury, between leucoreduced and non-leucoreduced group, the Pall BC1B leucocyte-depleting filter for blood cardioplegia has been shown to be effective in alleviating reperfusion injury in open-heart surgery patients.
Perfusion, Vol. 13, No. 3,
205-210 (1998)
DOI: 10.1177/026765919801300308

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
O. Warren, C. Alexiou, R. Massey, D. Leff, S. Purkayastha, J. Kinross, A. Darzi, and T. Athanasiou
The effects of various leukocyte filtration strategies in cardiac surgery
Eur. J. Cardiothorac. Surg.,
April 1, 2007;
31(4):
665 - 676.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. F. Salamonsen, J. Anderson, M. Anderson, M. Bailey, G. Magrin, and F. Rosenfeldt
Total Leukocyte Control for Elective Coronary Bypass Surgery Does Not Improve Short-Term Outcome
Ann. Thorac. Surg.,
June 1, 2005;
79(6):
2032 - 2038.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Samankatiwat, I. Samartzis, P. Lertsithichai, D. Stefanou, P. P Punjabi, K. M Taylor, and T. Gourlay
Leucocyte depletion in cardiopulmonary bypass: a comparison of four strategies
Perfusion,
March 1, 2003;
18(2):
95 - 105.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Martin, M. Krause, C. Benk, G. Lutter, M. Siegenthaler, and F. Beyersdorf
Blood cardioplegia filtration
Perfusion,
January 1, 2003;
18(1_suppl):
75 - 80.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
Poster presentation 2002
Perfusion,
January 1, 2003;
18(1_suppl):
89 - 90.
[PDF]
|
 |
|

|
 |

|
 |
 
G. A Ortolano, G. S Aldea, K. Lilly, P. O'Gara, J. D Alkon, F. Madera, T. Murad, C. P Altenbern, C. S Tritt, A. Capetandes, et al.
A review of leukofiltration in cardiac surgery: the time course of reperfusion injury may facilitate study design of anti-inflammatory effects
Perfusion,
March 1, 2002;
17(2_suppl):
53 - 62.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
G Matheis, M Scholz, A Simon, O. Dzemali, and A Moritz
Leukocyte filtration in cardiac surgery: a review
Perfusion,
September 1, 2001;
16(5):
361 - 370.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S J Morris
Leukocyte reduction in cardiovascular surgery
Perfusion,
September 1, 2001;
16(5):
371 - 380.
[PDF]
|
 |
|

|
 |

|
 |
 
G Matheis, M Scholz, A Simon, D Henrich, G Wimmer-Greinecker, and A Moritz
Timing of leukocyte filtration during cardiopulmonary bypass
Perfusion,
January 1, 2001;
16(1_suppl):
31 - 37.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
M. B Patel, K. S Kilgore, G. A Ortolano, C. L Gryboski, M. A Qureshi, P. Marcovitz, K. B Naylor, J. L Park, B. Wenz, N. Gikakis, et al.
Conditioned blood reperfusion during angioplasty (CoBRA) treatment of acute myocardial infarction
Perfusion,
January 1, 2001;
16(1_suppl):
39 - 49.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Roth, B. Kraus, T. Scheffold, O. Reuthebuch, W. P. Klovekorn, and E. P. Bauer
The effect of leukocyte-depleted blood cardioplegia in patients with severe left ventricular dysfunction: A randomized, double-blind study
J. Thorac. Cardiovasc. Surg.,
October 1, 2000;
120(4):
642 - 650.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P G Browning, M Pullan, M Jackson, and A Rashid
Leucocyte-depleted cardioplegia does not reduce reperfusion injury in hypothermic coronary artery bypass surgery
Perfusion,
September 1, 1999;
14(5):
371 - 377.
[Abstract]