Does heparin coating improve biocompatibility? A study on complement, blood cells and postoperative morbidity during cardiac surgery
Ali Belboul
Department of Thoracic and Cardiovascular Surgery, University of Gothenbourg, Sahlgrenska Hospital, Goteborg
Najib Al-Khaja
Department of Thoracic and Cardiovascular Surgery, University of Gothenbourg, Sahlgrenska Hospital, Goteborg
To evaluate whether the effect of heparin coating the extracorporeal circuit resulted in differences in patient outcome and haemostatic alteration, 24 patients undergoing elective, isolated coronary artery bypass were randomized prospectively to cardiopulmonary bypass (CPB) with heparin-coated circuits (group H, n = 12) or uncoated circuits (group C, n = 12). The technique of CPB, heparinization and its reversal were the same in both groups. We studied complement status (C3d, C3, C3d/C3, C4 and C-function), white blood cell counts with differentiation and the postoperative morbidity. The results confirmed that CPB activates complement and increases neutrophils in both the H and C groups. A significantly lower level of Ieucocytosis was seen in group H compared to the C group (p < 0.05). The complement function via the classical pathway (C-function), expressed as a percentage of the function of a reference serum pool (the values of normal sera were 75-125%), was significantly reduced in both heparin-coated and uncoated circuits (p < 0.05). There was no significant intergroup difference regarding C3, C3d/C3, C4 and C-function during the study period. A lower frequency of postoperative morbidity was present in the H group. We conclude that heparin-coated surfaces elicit less leucocytosis and decrease postoperative morbidity in patients undergoing cardiac surgery but do not cause a significant difference regarding activation of the complement system as reported by many other investigators.
Perfusion, Vol. 12, No. 6,
385-391 (1997)
DOI: 10.1177/026765919701200607

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
M. Ranucci, A. Balduini, A. Ditta, A. Boncilli, and S. Brozzi
A Systematic Review of Biocompatible Cardiopulmonary Bypass Circuits and Clinical Outcome
Ann. Thorac. Surg.,
April 1, 2009;
87(4):
1311 - 1319.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R Miraziz, L Hines, M Brouwer, R Steel, and P Klineberg
Bridging circuit for the resection of retroperitoneal sarcoma involving the aorta and the IVC - veno-venous to veno-arterial perfusion
Perfusion,
January 1, 2008;
23(1):
65 - 69.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
O. Mangoush, S. Purkayastha, S. Haj-Yahia, J. Kinross, M. Hayward, F. Bartolozzi, A. Darzi, and T. Athanasiou
Heparin-bonded circuits versus nonheparin-bonded circuits: an evaluation of their effect on clinical outcomes
Eur. J. Cardiothorac. Surg.,
June 1, 2007;
31(6):
1058 - 1069.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. P. Wendel, N. Hauser, F. Briquet, and G. Ziemer
Hemocompatibility of Medical Connectors with Biopassive or Bioactive Surface Coatings
J Biomater Appl,
July 1, 2002;
17(1):
5 - 17.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
H. A. Hennein
Inflammation After Cardiopulmonary Bypass: Therapy for the Postpump Syndrome
Seminars in Cardiothoracic and Vascular Anesthesia,
September 1, 2001;
5(3):
236 - 255.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Suhara, Y. Sawa, M. Nishimura, H. Oshiyama, K. Yokoyama, N. Saito, and H. Matsuda
Efficacy of a new coating material, PMEA, for cardiopulmonary bypass circuits in a porcine model
Ann. Thorac. Surg.,
May 1, 2001;
71(5):
1603 - 1608.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Belboul, C. Lofgren, C. Storm, and M. Jungbeck
Heparin-coated circuits reduce occult myocardial damage during CPB: a randomized, single blind clinical trial
Eur. J. Cardiothorac. Surg.,
May 1, 2000;
17(5):
580 - 586.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Svenmarker, E. Jansson, H. Stenlund, and K. G. Engstrom
Red blood cell trauma during cardiopulmonary bypass: narrow pore filterability versus free haemoglobin
Perfusion,
January 1, 2000;
15(1):
33 - 40.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
H. P. Wendel and G. Ziemer
Coating-techniques to improve the hemocompatibility of artificial devices used for extracorporeal circulation
Eur. J. Cardiothorac. Surg.,
September 1, 1999;
16(3):
342 - 350.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|
|