Sign In to gain access to subscriptions and/or personal tools.
|
Perfusion, Vol. 12, No. 5,
279-288 (1997)
DOI: 10.1177/026765919701200502
Australasian perfusion incident survey
Owen F Jenkins
Perfusion Department, St George Hospital, Sydney
Richard Morris
Perfusion Department, St George Hospital, Sydney
Judy M Simpson
Department of Public Health and Community Medicine, University of Sydney, Sydney
An anonymous postal survey about perfusion accidents, incidents and safety procedures was conducted retrospectively in all identifiable cardiac surgery units in Australia and New Zealand over an 18-month period from January 1994 to July 1995. Responses were received from 69% of all eligible perfusionists representing 39 of 42 operating units. The most frequent incidents reported were heater/cooler failure (43% of respondents), urgent return to bypass following circuit disposal (38%), air embolus in a circuit not reaching the patient (24%), accidental cannula displacement (28%), protamine-induced circuit clotting postbypass (20%), hospital power failure (31%) and oxygenator membrane leaks (24%). There were 11 serious injuries and 10 deaths reported, giving an overall rate of serious injury or death of one in 1300 cases. However, the perfusion-related injury rate was lower, at one injury or death per 2500 perfusions. The use of safety equipment was widespread with low-level alarms in use by all respondents, bubble detectors by 74%, arterial line filters by 82% and written or computerized checklists by 80%. While the rate of injury was lower than that reported in surveys done elsewhere, the rate of reporting of incidents was greater and this may reflect a changing attitude of perfusionists to accountability in the delivery of their services. The data suggest that the level of safety in perfusion in Australasia is high but that improved incident reporting may help to further improve practices.

CiteULike Connotea Del.icio.us Digg Reddit Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
J. W. Hammon
Extracorporeal Circulation: Perfusion System
Card. Surg. Adult,
January 1, 2008;
3(2008):
350 - 370.
[Full Text]
|
 |
|

|
 |

|
 |
 
M. P. Eason
Simulation Devices in Cardiothoracic and Vascular Anesthesia
Seminars in Cardiothoracic and Vascular Anesthesia,
December 1, 2005;
9(4):
309 - 323.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Svenmarker and M. Appelblad
Reporting of perfusion-related incidents: pitfalls and limitations
Perfusion,
September 1, 2005;
20(5):
243 - 248.
[PDF]
|
 |
|

|
 |

|
 |
 
D. A Palanzo
Perfusion safety: defining the problem
Perfusion,
July 1, 2005;
20(4):
195 - 203.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
W B Runciman and A F Merry
Crises in clinical care: an approach to management
Qual. Saf. Health Care,
June 1, 2005;
14(3):
156 - 163.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Kurusz and B. D Butler
Bubbles and bypass: an update
Perfusion,
January 1, 2004;
19(1_suppl):
S49 - S55.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Ginther Jr, R. Fillingham, B. Searles, and E. Darling
Departmental use of perfusion crisis management drills: 2002 survey results
Perfusion,
September 1, 2003;
18(5):
299 - 302.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
A. R. Fisher, M. Baker, M. Buffin, P. Campbell, S. Hansbro, S. Kennington, A. Lilley, and M. Whitehorne
Normal and abnormal trans-oxygenator pressure gradients during cardiopulmonary bypass
Perfusion,
January 1, 2003;
18(1):
25 - 30.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
E. A. Hessel II and L. H. Edmunds Jr.
Extracorporeal Circulation: Perfusion Systems
Card. Surg. Adult,
January 1, 2003;
2(2003):
317 - 338.
[Full Text]
|
 |
|

|
 |

|
 |
 
A. H Stammers and B. L Mejak
An update on perfusion safety: does the type of perfusion practice affect the rate of incidents related to cardiopulmonary bypass?
Perfusion,
May 1, 2001;
16(3):
189 - 198.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
B. L Mejak, A. Stammers, E. Rauch, S. Vang, and T. Viessman
A retrospective study on perfusion incidents and safety devices
Perfusion,
January 1, 2000;
15(1):
51 - 61.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Svenmarker, S. Haggmark, E. Jansson, R. Lindholm, M. Appelblad, and T. Aberg
Quality assurance in clinical perfusion
Eur. J. Cardiothorac. Surg.,
October 1, 1999;
14(4):
409 - 414.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F De Somer, L Foubert, E Schacht, and G Van Nooten
Nitric oxide donors attenuate increase in pressure drop across membrane oxygenators
Perfusion,
September 1, 1999;
14(5):
331 - 336.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Svenmarker and L. K von Segesser
The European Board of Cardiovascular Perfusion pre-bypass checklist
Perfusion,
May 1, 1999;
14(3):
165 - 166.
[PDF]
|
 |
|

|
 |

|
 |
 
A R Fisher
The incidence and cause of emergency oxygenator changeovers
Perfusion,
May 1, 1999;
14(3):
207 - 212.
[Abstract]
[PDF]
|
 |
|
|